AN-2000 COMPUTERIZED ANEMOMETRY SYSTEM
Technical Specifications
CTA CHANNEL MODULE
CTA BRIDGE:
Non Linear, Constant Temperature type: |
1:10 and 1:1 |
Sensor Resistance Range: |
0.3-200 Ohm |
Maximum Closed Loop Bandwidth: |
DC-120 kHz (up to 500 kHz. with option 04) |
Probe current: |
300 mA max |
Equivalent input noise: |
1.6nV/ |
Typical Output Noise: |
135�Vrms |
Stability - Typical Input Drift: Output |
0.4�V/�C |
SIGNAL CONDITIONER:
Output Voltage Range Amplifier Gain (256 discrete settings) Gain accuracy DC Offset Output Impendance Iutput Impendance Typical Input White Noise Frequency Range Typical Equivalent Input Drift (Gain 1) |
�12V 1 - 99.9 (Other gain values may be ordered). �0.2% +/-10V (in 5mV steps) 100Ohm 10KOhm 30nV / ![]() DC-200 KHz 160�V/�C |
LOW-PASS FILTER:
4'th order Butterworth Filter with 512 Cut-off frequencies. |
100Hz - 120KHz (Other frequency bands can be ordered). |
All filters have linear phase and are self calibrated to have the same delay on all channels. |
HIGH-PASS FILTER:
2 second order High-Pass filters, swichable on back panel. |
|
Cut-off frequencies | 1 Hz, 10 Hz |
CCA BRIDGE (OPTION 05):
Constant Current type (in addition to CTA mode) | |
Bridge current (in 256 equal steps) Setup accuracy Temperature drift |
0.5 - 50mA �0.5% 100ppm/�C |
CHANNEL A/D CONVERTER (OPTION 08):
Sampling method |
Simultaneous S&H 0-93 KHz , 0-160KHz. or 0-500KHz. optional internal or external Edge or Level Trigger 14 bit |
DIGITAL LINEARIZER (OPTION 09):
This option is available only if option 08 is installed.
Output resolution |
16 bit |
Output resolution is higher than Input resolution, in order to prevent errors caused by polynomial calculations. | |
All polynomial graphs and coefficients are calculated using floating point arithmetics. | |
All channels have analog "Top Of Bridge" and "Signal conditioner" outputs on BNC connectors. |
CONTROL UNIT
All analog parts are optically isolated from the digital parts and use separate power supplies.
DOT MATRIX DISPLAY:
Resolution |
256(W) x 128(H) pixels |
A/D CONVERTER
(For frequency response and scope only!)
Sampling Rate |
15 MHz 8 Bit |
All signals displayed are sampled by the control Unit A/D converter. | |
Acquired data which is sent to the parallel DMA port is sampled by the channel's A/D converter (14 bits). |
INTERFACES:
RS-232C serial interface |
Standard |
DMA parallel interface compatible with: DRV-11 connector Display bus |
Standard Standard |
Analog outputs | Top of bridge, buffered, amplified & filtered output |
Throughput rate External clock input External trigger input External Sync input (for Ensemble averaging) Operation voltage |
Up to 1.6 MegaWord/Sec(10MW optional) DC-90KHz,TTL level(DC-160 KHz. or DC-500 KHz. optional) TTL level TTL level 230V/115V 50/60Hz switchable. |
|