The
PMC-3000 is a measurement system, based on an analog data acquisition unit which
is
external to the computer and contained in a shielded box.
Each analog unit includes:
A highly accurate and
stable excitation for varying DC voltage levels for use in bridge type
transducers.
A low-cost solution for Force/Task
balance.
Menu driven PC software,
LABVIEW interface.
Connect to any Force/Pressure/Low-level
voltage sensor.
Built-in Signal conditioner.
Up to 12 channels, Simultaneous
sampling.
Three high-gain (up to 1000)
instrumentation amplifiers.
A low-pass filter
with a hardware selectable cut-off frequency, designed to cancel noise in
the
signal being sampled.
Three fast A/D converters
with simultaneous sample-and-hold providing a sampling rate of
40 kHz per channel.
The acquisition unit connects
to a PC-compatible computer via a digital interface card.
The 3-channel isolated data acquisition system connects with up to three analog
signal sources
which are amplified, conditioned and sampled.
Strain,
force & pressure measurement.
General purpose low-level
DC measurement.
Industrial process control.
Semiconductor temperature
sensors.
Watchdog / alarm level detector.
Computerized weighing
system.
The
PMC-3000 menu-driven "GAGE" software is used for data acquisition
and graphic display of the measured values. You can select any of the real-time
display modes:
Bar
display,
Time
Window scope,
Scrolling
scope, or
Fast
Courier Transform (FFT) of the input signals.

This software has many powerful functions, used for the analysis of data. It
can save the data and export the sampled data files to most of the Spread sheet
programs available to date. With this software, you can operate and calibrate
any transducer - much easier than before.
Sampled
measurements can be observed in real-time on user-defined scales. The displayed
graph
can be frozen at any time for further observation and analysis.
The digital interface card provides eight TTL digital I/O channels to simultaneous Input/Output signals from/to external instruments (relays, switches, lights, D/A units, buzzers, etc.). This Interface card also includes six digital input channels which can read the state of external events.
The external cards are optically isolated from the computer by industry standard opto-couplers capable of isolating voltages of up to 1000V. Optical isolation protects the computer from a sensor shorted to high voltage, and insulates the system from noise generated by the computer. The measurement system is floating in respect to the ground.
AMPLIFIER
Input
Channels: 3 to 12, differential and balanced to ground
Gain
Selection: Variable potentiometer adjustment (on front panel) with fine tuning
Gain
Temperature Coefficient: 50 ppm/deg.C
Input
Dynamic Range:
x 20 gain - ±150 mV
x 1000 gain - ±3 mV
Input
Impedance:
- Differential resistance 1 giga-Ohm
- Differential capacitance 10 pico-Farad
CMRR:
G=20 > 90 dB, DC to 60 Hz
G=100 > 100 dB, DC to 60 Hz
Nonlinearity:
0.001% max.
Gain
Bandwidth: DC to 10 kHz
Noise:
4 nV/root(Hz) @ 1 kHz, ±0.5 LSB at A/D input
Settling
Time: 30 uSec @ G=20, 75 uSec @ G=1000
DC
Offset: Continuously Variable adjustment with fine tuning
Maximum
Offset Range: ±20% of excitation
Temperature
Coefficient: 20 microVolt/deg.C
Power:
+5V, ±15V ext. power supply
Temperature
Range:
- Operating: 0 to +70 deg.C
- Storage: -10 to +100 deg.C
EXCITATION
Voltage
Range: 0 to 24 V adjustable, symmetrical about ground
Current:
200 mA max.
Protection:
Short-circuit protected
Temperature
Coefficient: 150 ppm/deg.C
A/D UNIT
The A/D Unit is optically isolated from the
computer.
Full Scale Input: ±3 Volts
Resolution:
12 Bits
Simultaneous
Sampling Rate:
40 kHz per channel (without optical isolation)
10 kHz per channel (isolated)
SNR:
72 dB
ALARM UNIT
The Alarm Unit has 16 preprogrammed hardware
tables which can be selected during measurement. These hardware tables can be
reprogrammed using the included software (EPROM programmer not included). Customized
tables can be ordered and programmed prior to delivery.
Level
Resolution: 2.9 mV
Response
Time: 200 nSec (from A/D readout)
Digital
Output: TTL compatible (0-5V)
Maximum
Output Current: 50 mA (sink)
The
PMC-3000 can be ordered with 3,6,9, or 12 input channels. The system includes
a Data acquisition & Amplifier Card, a PC compatible interface card, a floating
power supply and all the cabling needed.
The system runs on 110/220V, 50/60 Hz.
Alarm table may be standard (S) or customized for your requirements (C).
EXAMPLE:
PMC - 3000 - 6 - S
Where:
6
= No. of Channels
S = Alarm Table, standard
For more information about our Amplifiers and Signal Conditioner products , please click here.
For more information please contact us at:
Head Office: 33 Hayetzira
St., Ramat-Gan 52521, ISRAEL. Tel: 972-3-5756327, Fax: 972-3-5756326
USA office: 9 Blossom drive, Kennett Square, PA19348.
Tel: (302)478-2881 Fax: (610)444-5544
Please note: New USA address and new fax number.
E-mail:
info@lab-systems.com
© Copyright 2000, A.A.Lab Systems Ltd.